[Home] [Databases] [World Law] [Multidatabase Search] [Help] [Feedback] | ||
United Kingdom Statutory Instruments |
||
You are here: BAILII >> Databases >> United Kingdom Statutory Instruments >> The Measuring Instruments (Non-Prescribed Instruments) Regulations 2006 No. 1270 URL: http://www.bailii.org/uk/legis/num_reg/2006/20061270.html |
[New search] [Help]
Made | 28th April 2006 | ||
Laid before Parliament | 8th May 2006 | ||
Coming into force | |||
Regulations 1, 2, 4 6 and 7 and Part 1 of Schedule 2 | 30th May 2006 | ||
Remaining regulations | 30th October 2006 |
1. | Citation and commencement |
2. | Interpretation |
3. | Application |
4. | Designation of notified bodies |
5. | Functions of notified bodies |
6. | Provisions supplemental to regulation 4 |
7. | Variation and termination of designations |
8. | Fees |
9. | Compliance with the essential requirements |
10. | Conformity assessment procedures |
11. | Marking and identification requirements |
12. | Conformity with other directives |
SCHEDULE 1— |
PART 1— | ESSENTIAL REQUIREMENTS |
PART 2— | INSTRUMENT SPECIFIC REQUIREMENTS |
SCHEDULE 2— | NOTIFIED BODIES |
PART 1— | NOTIFIED BODY CRITERIA |
PART 2— | FUNCTIONS |
SCHEDULE 3— | TECHNICAL DOCUMENTATION |
SCHEDULE 4— | MARKING AND INSCRIPTIONS |
and who has been notified to the Commission and the other member States pursuant to Article 11.1;
makes up a measuring instrument;
(2) In these Regulations a reference to a member State includes Norway, Iceland and Liechtenstein[6].
(3) References in these Regulations to an Article, Annex or a part of an Annex are references to an Article, Annex, or a part of an Annex to the Directive.
Application
3.
—(1) These Regulations apply to the following measuring instruments—
where such measuring instruments must satisfy the essential requirements in relation to legal metrological control in order to be placed on the market or put into use in another member State on or after 30th October 2006.
(2) For the purposes of this regulation "legal metrological control" means the control of the measurement tasks intended for the field of application of a measuring instrument for reasons of public health, public safety, public order, protection of the environment, levying of taxes and duties, protection of consumers and fair trading.
Designation of notified bodies
4.
—(1) The designating authority may, on the application of a person, designate that person to be a notified body.
(2) The designating authority shall not make a designation under paragraph (1) unless it is satisfied that the person meets the notified body criteria.
(3) A person who meets the criteria laid down in a national standard shall be presumed to meet that part of the notified body criteria which corresponds to the criteria in the national standard.
(4) A designation under paragraph (1)—
(c) subject to regulation 7, may last for such period as may be specified in the designation;
(d) shall specify the conformity assessment procedures and specific tasks (which may be framed by reference to any circumstances) which the person has been designated to carry out; and
(e) may include a requirement to publish from time to time the scale of fees which the person charges pursuant to regulation 8 or such information about the basis of calculation of such fees as may be specified.
(5) In exercising the power conferred on it by paragraph (1), the designating authority may (in addition to the matters of which it is required to satisfy itself pursuant to paragraph (2)) have regard to any matter appearing to it to be relevant.
(6) The Gas and Electricity Markets Authority and the Northern Ireland Authority for Energy Regulation shall be entitled—
(7) For the purposes of paragraph (3), "national standard" means a standard applicable to the designation of notified bodies—
Functions of notified bodies
5.
A notified body shall carry out the functions set out in Part 2 of Schedule 2.
Provisions supplemental to regulation 4
6.
—(1) The Secretary of State shall, from time to time, publish a list of notified bodies indicating the description of any measuring instrument in respect of which a notified body other than the Secretary of State is designated; and such a list may include information concerning any condition to which the designation of any notified body other than the Secretary of State is subject.
(2) The designating authority shall, from time to time, carry out an inspection of notified bodies which it has designated under regulation 4 with a view to verifying that the notified body—
but, unless it appears to it that there are circumstances which make it necessary or expedient to do so, it shall not carry out an inspection within two years from the date of designation under regulation 4, or of any later inspection under this paragraph.
Variation and termination of designations
7.
—(1) The designating authority may vary a designation made by it under regulation 4 if—
(2) The designating authority may terminate a designation made by it under regulation 4—
(3) Where the designating authority is minded to—
it shall—
(4) If a designation is terminated under paragraph (2), the designating authority may—
Fees
8.
—(1) A notified body other than the Secretary of State may charge such fees in connection with, or incidental to, the carrying out of the conformity assessment procedures or specific tasks as it may determine.
(2) The fees referred to in paragraph (1) shall not exceed the following—
(3) The power in paragraph (1) includes the power to require the payment of fees or a reasonable estimate of such fees in advance of carrying out the work requested by the applicant.
(4) Where any fees payable to a notified body pursuant to this regulation remain unpaid 28 days after either the work has been completed or payment of the fees has been requested in writing, whichever is the later, that notified body may by 14 days' notice in writing provide that, unless the fees are paid before the expiry of the notice, the certificate or notification appropriate to the relevant conformity assessment procedure may be suspended until payment of the fees has been received.
Compliance with the essential requirements
9.
—(1) Where a measuring instrument must be compliant with the essential requirements in order to be placed on the market or put into use in another member State, the manufacturer may demonstrate that the measuring instrument is compliant with the essential requirements by—
and selecting and following one of the conformity assessment procedures.
(2) A measuring instrument which is compliant with the relevant national standard or the relevant normative document shall be presumed to be compliant with the essential requirements.
(3) Where the instrument is compliant only in part with the relevant national standard or relevant normative document, it shall be presumed to be compliant only with that part of the essential requirements which corresponds to the element of the relevant national standard or relevant normative document with which the instrument is compliant.
Conformity assessment procedures
10.
—(1) The conformity assessment procedures are the procedures as follows—
(ii) for electromechanical instruments—
(iii) for electronic systems or systems containing software—
(b) for capacity serving measures—
(c) for dimensional measuring instruments—
(ii) for electronic instruments or instruments containing software—
(d) for heat meters—
(e) for material measures of length—
(f) for measuring systems of the continuous and dynamic measurement of quantities of liquid other than water—
(g) for volume conversion devices—
(h) for water meters—
(2) The manufacturer or his authorised representative shall provide to the notified body carrying out the relevant conformity assessment procedure the technical documentation set out in Schedule 3.
(3) In this regulation—
Marking and identification requirements
11.
—(1) Where a measuring instrument is compliant with the essential requirements—
(2) Any other marking may be affixed to the measuring instrument provided that the visibility and legibility of the CE marking, the M marking and the identification number of the notified body are not reduced.
(3) For the purposes of paragraph (1)—
Conformity with other directives
12.
—(1) Where a measuring instrument falls within the scope of other directives which provide for the affixing of the CE marking, the affixing of the CE marking under these Regulations shall indicate that the instrument is also presumed to be compliant with the requirements of those other directives.
(2) Where paragraph (1) applies, the publication reference of such other directives in the Official Journal of the European Union must be given in the documents, notices or instructions required to accompany the instrument.
Barry Gardiner
Parliamentary Under Secretary of State for Competitiveness Department of Trade and Industry
28th April 2006
Allowable Errors
3.
—(1) Under rated operating conditions and in the absence of a disturbance, the error of measurement shall not exceed the maximum permissible error (MPE) value as set out in the appropriate instrument-specific requirements.
(2) Unless stated otherwise in the instrument-specific provisions, MPE is expressed as a bilateral value of the deviation from the true measurement value.
(3) Under rated operating conditions and in the presence of a disturbance, the performance requirement shall be as set out in the appropriate instrument specific requirements.
(4) Where the instrument is intended to be used in a specified permanent continuous electromagnetic field the permitted performance during the radiated electromagnetic field-amplitude modulated test shall be within MPE.
(5) The manufacturer shall specify the climatic, mechanical and electromagnetic environments in which the instrument is intended to be used, power supply and other influence quantities likely to affect its accuracy, taking account of the instrument-specific requirements.
Temperature Limits | ||||
Upper temperature limit | 30°C | 40°C | 55°C | 70°C |
Lower temperature limit | 5°C | -10°C | -25°C | -40°C |
(c) Electromagnetic environments are classified into classes E1, E2 or E3, unless otherwise laid down in the appropriate instrument-specific provisions in Part 2 of this Schedule—
(ii) The following influence quantities shall be considered in relation with electromagnetic environments—
(6) Other influence quantities to be considered, where appropriate, are—
(7) When carrying out the tests as envisaged in this Schedule, the following paragraphs apply—
(b) Ambient humidity—
Reproducibility
4.
The application of the same measurand in a different location or by a different user, all other conditions being the same, shall result in the close agreement of successive measurements. The difference between the measurement results shall be small when compared with the MPE.
Repeatability
5.
The application of the same measurand under the same conditions of measurement shall result in the close agreement of successive measurements. The difference between the measurement results shall be small when compared with the MPE.
Discrimination and Sensitivity
6.
A measuring instrument shall be sufficiently sensitive and the discrimination threshold shall be sufficiently low for the intended measurement task.
Durability
7.
A measuring instrument shall be designed to maintain an adequate stability of its metrological characteristics over a period of time estimated by the manufacturer, provided that it is properly installed, maintained and used according to the manufacturer's instruction when in the environmental conditions for which it is intended.
Reliability
8.
A measuring instrument shall be designed to reduce as far as possible the effect of a defect that would lead to an inaccurate measurement result, unless the presence of such a defect is obvious.
Suitability
9.
—(1) A measuring instrument shall have no feature likely to facilitate fraudulent use, whereas possibilities for unintentional misuse shall be minimal.
(2) A measuring instrument shall be suitable for its intended use taking account of the practical working conditions and shall not require unreasonable demands of the user in order to obtain a correct measurement result.
(3) The errors of a utility measuring instrument at flows or currents outside the controlled range shall not be unduly biased.
(4) Where a measuring instrument is designed for the measurement of values of the measurand that are constant over time, the measuring instrument shall be insensitive to small fluctuations of the value of the measurand, or shall take appropriate action.
(5) A measuring instrument shall be robust and its materials of construction shall be suitable for the conditions in which it is intended to be used.
(6) A measuring instrument shall be designed so as to allow the control of the measuring tasks after the instrument has been placed on the market and put into use. If necessary, special equipment or software for this control shall be part of the instrument. The test procedure shall be described in the operation manual.
(7) When a measuring instrument has associated software which provides other functions besides the measuring function, the software that is critical for the metrological characteristics shall be identifiable and shall not be inadmissibly influenced by the associated software.
Protection against corruption
10.
—(1) The metrological characteristics of a measuring instrument shall not be influenced in any inadmissible way by the connection to it of another device, by any feature of the connected device itself or by any remote device that communicates with the measuring instrument.
(2) A hardware component that is critical for metrological characteristics shall be designed so that it can be secured. Security measures foreseen shall provide for evidence of an intervention.
(3) Software that is critical for metrological characteristics shall be identified as such and shall be secured.
(4) Software identification shall be easily provided by the measuring instrument.
(5) Evidence of a software intervention shall be available for a reasonable period of time.
(6) Measurement data, software that is critical for measurement characteristics and metrologically important parameters stored or transmitted shall be adequately protected against accidental or intentional corruption.
(7) For utility measuring instruments the display of the total quantity supplied or the displays from which the total quantity supplied can be derived, whole or partial reference to which is the basis for payment, shall not be able to be reset during use.
Information to be borne by and to accompany the instrument
11.
—(1) A measuring instrument shall bear the following inscriptions:
plus, when applicable—
(2) An instrument of dimensions too small or of too sensitive a composition to allow it to bear the relevant information shall have its packaging, if any, and the accompanying documents required by the provisions of these Regulations suitably marked.
(3) The instrument shall be accompanied by information on its operation, unless the simplicity of the measuring instrument makes this unnecessary. Information shall be easily understandable and shall include where relevant—
(4) Groups of identical measuring instruments used in the same location or used for utility measurements do not necessarily require individual instruction manuals.
(5) Unless specified otherwise in an instrument-specific annex, the scale interval for a measured value shall be in the form 1 x 10n, 2 x 10n or 5 x 10n, where n is any integer or zero. The unit of measurement or its symbol shall be shown close to the numerical value.
(6) A material measure shall be marked with a nominal value or a scale, accompanied by the unit of measurement used.
(7) The units of measurement used and their symbols shall be in accordance with the provisions of Community legislation on units of measurement and their symbols.
(8) All marks and inscriptions required under any requirement shall be clear, non-erasable, unambiguous and non-transferable.
Indication of result
12.
—(1) Indication of the result shall be by means of a display or hard copy.
(2) The indication of any result shall be clear and unambiguous and accompanied by such marks and inscriptions necessary to inform the user of the significance of the result. Easy reading of the presented result shall be permitted under normal conditions of use. Additional indications may be shown provided they cannot be confused with the metrologically controlled indications.
(3) In the case of hard copy the print or record shall also be easily legible and non-erasable.
(4) A measuring instrument for direct sales trading transactions shall be designed to present the measurement result to both parties in the transaction when installed as intended. When critical in case of direct sales, any ticket provided to the consumer by an ancillary device not complying with the appropriate requirements of these Regulations shall bear an appropriate restrictive information.
(5) Whether or not a measuring instrument intended for utility measurement purposes can be remotely read it shall in any case be fitted with a metrologically controlled display accessible without tools to the consumer. The reading of this display is the measurement result that serves as the basis for the price to pay.
Further processing of data to conclude the trading transaction
13.
—(1) A measuring instrument other than a utility measuring instrument shall record by a durable means the measurement result accompanied by information to identify the particular transaction, when:
(2) Additionally, a durable proof of the measurement result and the information to identify the transaction shall be available on request at the time the measurement is concluded.
Conformity evaluation
14.
A measuring instrument shall be designed so as to allow ready evaluation of its conformity with the appropriate requirements of these Regulations.
Rated Operating Conditions
17.
The manufacturer shall specify the rated operating conditions for the instrument, in particular—
(1) The flowrate range of the water.
(b) For 5 years from the date of entry into force of these Regulations the ratio Q2/Q1may be: 1.5, 2.5, 4 or 6.3.
(2) The temperature range of the water.
(b) The meter may be designed to operate over both ranges.
(3) The relative pressure range of the water, the range being 0.3 bar to at least 10 bar at Q3.
(4) For the power supply: the nominal value of the AC voltage supply and/or the limits of DC supply.
MPE
18.
—(1) The MPE, positive or negative, on volumes delivered at flowrates between the transitional flowrate (Q2) (included) and the overload flowrate (Q4) is—
(2) The MPE, positive or negative, on volumes delivered at flowrates between the minimum flowrate (Q1) and the transitional flowrate (Q2) (excluded) is 5% for water having any temperature.
Permissible Effect of Disturbances
19.
—(1) Electromagnetic immunity
(b) After undergoing an electromagnetic disturbance the water meter shall—
(c) The critical change value is the smaller of the two following values—
(2) Durability
(ii) The error of indication for the volume metered after the durability test shall not exceed—
Suitability
20.
—(1) The meter shall be able to be installed to operate in any position unless clearly marked otherwise.
(2) The manufacturer shall specify whether the meter is designed to measure reverse flow. In such a case, the reverse flow volume shall either be subtracted from the cumulated volume or shall be separately recorded. The same MPE shall apply to both forward and reverse flow.
(3) Water meters not designed to measure reverse flow shall either prevent reverse flow or shall withstand an accidental reverse flow without any deterioration or change in metrological properties.
Units of Measurement
21.
Metered volume shall be displayed in cubic metres.
Putting into Use
22.
The member State shall ensure that the requirements under paragraphs 17(1), (2) and (3) are determined by the distributor or the person legally designated for installing the meter, so that the meter is appropriate for the accurate measurement of consumption that is foreseen or foreseeable.
Rated operating conditions
25.
The values of the rated operating conditions shall be specified by the manufacturer as follows—
Accuracy classes
26.
The following accuracy classes are defined for heat meters:
MPE applicable to complete heat meters
27.
The maximum permissible relative errors applicable to a complete heat meter, expressed in percent of the true value of each accuracy class, are—
Permissible influences of electromagnetic disturbances
28.
—(1) The instrument shall not be influenced by static magnetic fields and by electromagnetic fields at mains frequency.
(2) The influence of an electromagnetic disturbance shall be such that the change in the measurement result is not greater than the critical change value as laid down in sub-paragraph (3) or the indication of the measurement result is such that it cannot be interpreted as a valid result.
(3) The critical change value for a complete heat meter is equal to the absolute value of the MPE applicable to that heat meter (see paragraph 27).
Durability
29.
—(1) After an appropriate test, taking into account the period of time estimated by the manufacturer, has been performed, the following criteria shall be satisfied—
Inscriptions on a heat meter
30.
The following must be inscribed on the meter—
Sub-assemblies
31.
The provisions for sub-assemblies may apply to sub-assemblies manufactured by the same or different manufacturers. Where a heat meter consists of sub-assemblies, the essential requirements for the heat meter apply to the sub-assemblies as relevant. In addition, the following apply—
where the error Ef relates the indicated value to the true value of the relationship between flow sensor output signal and the mass or the volume.
(b) The relative MPE of the temperature sensor pair, expressed in % —
Et = (0.5 + 3 Δθmin /Δθ),
where the error Et relates the indicated value to the true value of the relationship between temperature sensor pair output and temperature difference.
(c) The relative MPE of the calculator, expressed in % —
Ec = (0.5 + Δθmin /Δθ),
where the error Ec relates the value of the heat indicated to the true value of the heat.
(d) The critical change value for a sub-assembly of a heat meter is equal to the respective absolute value of the MPE applicable to the sub-assembly (see sub-paragraphs (a), (b) or (c)).
(e) Inscriptions on the sub-assemblies—
(ii) Temperature sensor pair—
(iii) Calculator—
Putting into use
32.
—(1) Where a member State imposes measurement of residential use, it shall allow such measurement to be performed by means of any Class 3 meter.
(2) Where a member State imposes measurement of commercial and/or light industrial use, it is authorised to require any Class 2 meter.
(3) As regards the requirements under paragraph 25(a) to (d), member States shall ensure that the properties be determined by the distributor or the person legally designated for installing the meter, so that the meter is appropriate for the accurate measurement of consumption that is foreseen or foreseeable.
Rated operating conditions
35.
—(1) The manufacturer shall specify the rated operating conditions for the instrument, in particular—
(2) The flowrate range.
The flowrate range is subject to the following conditions—
Specific measuring system | Characteristic of liquid | Minimum ratio of Qmax : Qmin |
Fuel dispenser | Not liquefied gases | 10:1 |
Liquefied gases | 5:1 | |
Measuring system | Cryogenic liquids | 5:1 |
Measuring systems on pipeline and systems for loading ships | All liquids | Suitable for use |
All other measuring systems | All liquids | 4:1 |
(4) The normal value of the AC voltage supply and/or limits of the DC voltage supply.
(5) The base conditions for converted values.
Note: Sub-paragraph (4) is without prejudice to the member States' obligations to require use of a temperature of either 15°C in accordance with Article 3(1) of Council Directive 92/81/EEC of 19th October 1992 on the harmonisation of the structures of excise duties on mineral oils[17] or, for heavy fuel oils, LPG and methane, another temperature pursuant to Article 3(2) of that Directive.
Accuracy classification and maximum permissible errors (MPEs)
36.
—(1) For quantities equal to or greater than 2 litres the MPE on indications is:
Table 3
Accuracy class
0.3
0.5
1.0
1.5
2.5
Measuring systems (A)
0.3%
0.5%
1.0%
1.5%
2.5%
Meters (B)
0.2%
0.3%
0.6%
1.0%
1.5%
(2) For quantities less than two litres the MPE on indications is:
Table 4
Measured volume V
MPE
V < 0.1 L
4 x value in Table 3, applied to 0.1 L
0.1 L ≤ V < 0.2 L
4 x value in Table 3
0.2 L ≤ V < 0.4 L
2 x value in Table 3, applied to 0.4 L
0.4 L ≤ V < 1 L
2 x value in Table 3
1 L ≤ V < 2 L
Value in Table 3, applied to 2 L
(3) However, no matter what the measured quantity may be, the magnitude of the MPE is given by the greater of the following two values—
(a) the absolute value of the MPE given in Table 3 or Table 4;
(b) the absolute value of the MPE for the minimum measured quantity (Emin ).
(4)
(b) For minimum measured quantities of less than two litres, condition 1 applies and Emin is twice the value specified in Table 4, and related to line A of Table 3.
(5) Converted indication—
(6) Conversion devices—
MPE on measurements | Accuracy classes of the measuring system | ||||
0.3 | 0.5 | 1.0 | 1.5 | 2.5 | |
Temperature | ± 0.3 °C | ± 0.5 °C | ± 1.0 °C | ||
Pressure |
Less than 1 MPa: ± 50 kPa From 1 to 4 MPa: ± 5 % Over 4 MPa: ± 200 kPa |
||||
Density | ± 1kg/m³ | ± 2kg/m³ | ± 5kg/m³ |
These values apply to the indication of the characteristic quantities of the liquid displayed by the conversion device.
The MPE for the calculation of each characteristic quantity of the liquid, positive or negative, is equal to two fifths of the value fixed in (ii).
(7) The requirement (i) in sub-paragraph (6) applies to any calculation, not only conversion.
Maximum permissible effect of disturbances
37.
—(1) The effect of an electromagnetic disturbance on a measuring system shall be one of the following—
(2) The critical change value is the greater of MPE/5 for a particular measured quantity or Emin.
Durability
38.
After an appropriate test, taking into account the period of time estimated by the manufacturer, has been performed, the following criterion shall be satisfied—
Suitability
39.
—(1) For any measured quantity relating to the same measurement, the indications provided by various devices shall not deviate one from another by more than one scale interval where devices have the same scale interval. In the case where the devices have different scale intervals, the deviation shall not be more than that of the greatest scale interval.
(2) In the case of a self-service arrangement the scale intervals of the main indicating device on the measuring system and the scale intervals of the self-service device shall be the same and results of measurement shall not deviate one from another.
(3) It shall not be possible to divert the measured quantity in normal conditions of use unless it is readily apparent.
(4) Any percentage of air or gas not easily detectable in the liquid shall not lead to a variation of error greater than—
The allowed variation shall never be smaller than 1% of MMQ. This value applies in the case of air or gas pockets.
(5) Instruments for direct sales
(6) Fuel Dispensers
Power supply failure
40.
A measuring system shall either be provided with an emergency power supply device that will safeguard all measuring functions during the failure of the main power supply device or be equipped with means to save and display the data present in order to permit the conclusion of the transaction in progress and with means to stop the flow at the moment of the failure of the main power supply device.
41.
Putting into use
Table 6
Accuracy Class
Types of Measuring system
0.3
Measuring systems on pipeline
0.5
All measuring systems if not differently stated elsewhere in this Table, in particular:
— fuel dispensers (not for liquified gases),
— measuring system on road tankers for liquids of low viscosity (< 20 mPa.s),
— measuring systems for (un)loading ships and rail and road tankers(¹),
— measuring systems for milk,
— measuring systems for refuelling aircraft.
1.0
Measuring system for liquefied gases under pressure measured at a temperature equal to or above -10°C
Measuring systems normally in class 0.3 or 0.5 but used for liquids—
— whose temperature is less than -10°C or greater than 50°C;
— whose dynamic viscosity is higher than 1000 mPa.s;
— whose maximum volumetric flowrate is not higher than 20 L/h.
1.5
Measuring systems for liquefied carbon dioxide
Measuring systems for liquefied gases under pressure measured at a temperature below –10ºC (other than cryogenic liquids)
2.5
Measuring systems for cryogenic liquids (temperature below –153ºC)
(¹) However, Member States may require measuring systems of accuracy class 0.3 or 0.5 when used for the levying of duties on mineral oils when (un)loading ships and rail and road tankers.
Note: However, the manufacturer may specify a better accuracy for a certain type of measuring system
Units of measurement
42.
The metered quantity shall be displayed in millilitres, cubic centimetres, litres, cubic metres, grams, kilogram or tonnes.
(b) For the electrical supply influence quantities—
(c) For the mechanical and climatic influence quantities—
(d) For other influence quantities (if applicable)—
Suitability
45.
—(1) Means shall be provided to limit the effects of tilt, loading and rate of operation such that maximum permissible errors (MPEs) are not exceeded in normal operation.
(2) Adequate material handling facilities shall be provided to enable the instrument to respect the MPEs during normal operation.
(3) Any operation control interface shall be clear and effective.
(4) The integrity of the display (where present) shall be verifiable by the operator.
(5) Adequate zero setting capability shall be provided to enable the instrument to respect the MPEs during normal operation.
(6) Any result outside the measurement range shall be identified as such, where a printout is possible.
Automatic Catchweighers
46.
—(1) In addition to the requirements set out in paragraphs 44 and 45, the following specific requirements shall apply in relation to automatic catchweighers.
(2) Definitions—
(3) Accuracy Classes
which shall be specified by the manufacturer.
(c) Category X applies to instruments used to check pre-packages made up in accordance with the requirements of Council Directive 75/106/EEC of 19 December 1974 on the approximation of the laws of the member States relating to the making-up by volume of certain pre-packaged liquids[18] and of Council Directive 76/211/EEC of 20th January 1976 on the approximation of the laws of the member States relating to the making-up by weight or by volume of certain pre-packaged products[19] applicable to pre-packages.
(d) The accuracy classes are supplemented by a factor (x) that qualifies the maximum permissible standard deviation as specified in sub-paragraph (3)(b).
(e) The manufacturer shall specify the factor (x), where (x) shall be ≤ 2 and in the form 1 x 10k, 2 x 10k or 5 x 10k, where k is a negative whole number or zero.
(f) Category Y applies to all other automatic catchweighers.
(4) MPE
Net Load (m) in verification scale intervals (c) | Maximum permissible mean error | Maximum permissible error | |||||||
XI | Y(1) | XII | Y(II) | XIII | Y(a) | XIV | Y(b) | X | Y |
0 < m ≤ 50 000 | 0 < m ≤ 5 000 | 0 < m ≤ 500 | 0 < m ≤ 50 | ± 0.5 e | ±1 e | ||||
50 000 < m ≤ 200 000 | 5 000 < m ≤ 20 000 | 500 < m ≤ 2 000 | 50 < m ≤ 200 | ± 1.0 e | ±1.5 e | ||||
200 000 < m | 20 000 < m ≤ 100 000 | 2 000 < m ≤ 10 000 | 200 < m ≤ 1 000 | ± 1.5 e | ± 2 e |
Net Load (m) | Maximum permissible standard deviation for class X(1) |
m ≤ 50g | 0.48% |
50 g < m ≤ 100 g | 0.24 g |
100 g < m ≤ 200 g | 0.24 % |
200 g < m ≤ 300 g | 0.48 g |
300 g < m ≤ 500 g | 0.16 % |
500g < m ≤ 1 000 g | 0.8 g |
1 000 g < m ≤ 10 000 g | 0.08 % |
10 000 g < m ≤ 15 000 g | 8 g |
15 000 g < m | 0.053 % |
For class XI and XII (x) shall be less than 1. For class XIII (x) shall be not greater than 1. For class XIV (x) shall be greater than 1. |
Accuracy classes | Verification scale interval | Number of verification scale intervals n= Max/e | ||
Minimum | Maximum | |||
XI | Y(I) | 0.001 g ≤ e | 50 000 | – |
XII | Y(II) |
0.001 g ≤ e ≤ 0.05 g |
100 | 100 000 |
0.1 g ≤ e | 5 000 | 100 000 | ||
XIII | Y(a) | 0.1 g ≤ e ≤ 2 g | 100 | 10 000 |
5 g ≤ e | 500 | 10 000 | ||
XIV | Y(b) | 5 g ≤ e | 100 | 1 000 |
Accuracy classes | Verification scale interval | Number of verification scale intervals n=Max/e | ||
Minimum value(¹) n = Maxi/e (i+1) |
Maximum n=Maxi/ei |
|||
XI | Y(I) | 0.001 g ≤ ei | 50 000 | – |
XII | Y(II) |
0.001 g ≤ ei 0.05 g |
5 000 | 100 000 |
0.1 g ≤ ei | 5 000 | 100 000 | ||
XIII | Y(a) | 0.1 g ≤ ei | 500 | 10 000 |
XIV | Y(b) | 5 g ≤ ei | 50 | 1 000 |
(¹) For i = r the corresponding column of Table 9 applies with e replaced by er.
Where:
(5) Measurement Range
In specifying the measurement range for class Y instruments the manufacturer shall take account that the minimum capacity shall not be less than:
(6) Dynamic setting
(7) Performance under influence factors and electromagnetic disturbances
(ii) For category Y instruments—
(b) The critical change value due to a disturbance is one verification scale interval.
(c) Temperature range
Automatic gravimetric filling instruments
47.
—(1) In addition to the requirements set out in paragraphs 44 and 45, the following specific requirements shall apply in relation to automatic gravimetric filling instruments.
(2) Accuracy classes
(3) MPE
(b) Deviation from average fill
Table 11
Value of the mass of the fills (m) in grams
Maximum permissible deviation of each fill from the average for class X(1)
m ≤ 50
7.2%
50 < m ≤ 100
3.6 grams
100 < m ≤ 200
3.6 %
200 < m ≤ 300
7.2 grams
300 < m ≤ 500
2.4 %
500 < m ≤ 1 000
12 grams
1 000 < m ≤ 10 000
1.2 %
10 000 < m ≤ 15 000
120 grams
15 000 < m
0.8 %
Note: The calculated deviation of each fill from the average may be adjusted to take account of the effect of material particle size
(c) Error relative to pre-set value (setting errors)
For automatic gravimetric filling instruments where it is possible to pre-set a fill weight, the maximum difference between the pre-set value and the average mass of the fills shall not exceed 0.312 of the maximum permissible deviation of each fill from the average, as specified in Table 11.
(4) Performance Under Influence Factor and Electromagnetic Disturbance
Discontinuous Totalisers
48.
—(1) In addition to the requirements set out in paragraphs 44 and 45, the following specific requirements shall apply in relation to discontinuous totalisers.
(2) Accuracy Classes
Discontinuous totalisers are divided into four accuracy classes, as follows—
(3) MPE
Table 12
Accuracy class
MPE of Totalised load
0.2
± %
0.5
± %
1
± %
2
± %
(4) Totalisation scale interval
The totalisation scale interval (dt) shall be in the range:
0.01 % Max ≤ dt ≤ % Max
(5) Minimum Totalised Load (Σmin)
The minimum totalised load (Σmin) shall be not less than the load at which the MPE is equal to the totalisation scale interval (dt) and not less than the minimum load as specified by the manufacturer.
(6) Zero Setting
Instruments that do not tare weigh after each discharge shall have a zero setting device. Automatic operation shall be inhibited if zero indication varies by—
(7) Operator Interface
Operator adjustments and reset function shall be inhibited during automatic operation.
(8) Printout
On instruments equipped with a printing device, the reset of the total shall be inhibited until the total is printed. The printout of the total shall occur if automatic operation is interrupted.
(9) Performance under influence factors and electromagnetic disturbances
Load (m) in totalisation scale intervals dt | MPE |
0 < m ≤ 500 | ± dt |
500 < m ≤ 2 000 | ± dt |
2 000 < m ≤ 10 000 | ± dt |
Continuous Totalisers
49.
—(1) In addition to the requirements set out in paragraphs 44 and 45, the following specific requirements shall apply in relation to continuous totalisers.
(2) Accuracy Classes
Continuous totalisers are divided into three accuracy classes, as follows—
(3) Measurement Range
(4) MPE
Table 14
Accuracy class
MPE for totalised load
0.5
± 0.25%
1
± 0.5%
2
± 1.0%
(5) Speed of the Belt
The speed of the belt shall be specified by the manufacturer. For single-speed beltweighers, and variable-speed beltweighers having a manual speed setting control, the speed shall not vary by more than 5% of the nominal value. The product shall not have a different speed than the speed of the belt.
(6) General Totalisation Device
It shall not be possible to reset the general totalisation device to zero.
(7) Performance under influence factor and electromagnetic disturbance
(a) The MPE due to influence factor, for a load not less than Σmin, shall be 0.7 times the appropriate value specified in Table 14, rounded to the nearest totalisation scale interval (d).
(b) The critical change value due to a disturbance shall be 0.7 times the appropriate value specified in Table 14, for a load equal to Σmin, for the designated class of beltweigher, rounded up to the next higher totalisation scale interval (d).
Automatic Rail-weighbridges
50.
—(1) In addition to the requirements set out in paragraphs 44 and 45, the following specific requirements shall apply in relation to automatic rail-weighbridges.
(2) Accuracy classes
Automatic rail-weighbridges are divided into four accuracy classes as follows:
(3) MPE
Accuracy class | MPE |
0.2 | ± 0.1 % |
0.5 | ± 0.25 % |
1 | ± 0.5 % |
2 | ± 1.0 % |
(c) The MPEs for the weight of train weighing-in-motion shall be one of the following values, whichever is the greatest:
(d) When weighing coupled wagons; the errors of not more than 10% of the weighing results taken from one or more passes of the train may exceed the appropriate MPE given in paragraph 3(b), but shall not exceed twice the MPE.
(4) Scale interval (d)
The relationship between the accuracy class and the scale interval shall be as specified in Table 16.
Table 16
Accuracy class
Scale interval (d)
0.2
d ≤ 50 kg
0.5
d ≤ 100 kg
1
d ≤ 200 kg
5
d ≤ 500 kg
(5) Measurement range
(a) The minimum capacity shall not be less than 1 t, and not greater than the value of the result of the minimum wagon weight divided by the number of partial weighings.
(b) The minimum wagon weight shall not be less than 50 d.
(6) Performance under influence factor and electromagnetic disturbance
Load (m) in verification scale intervals (d) | MPE |
0 < m ≤ 500 | ± 0.5 d |
500 < m ≤ 2 000 | ± 1.0 d |
2 000 < m ≤ 10 000 | ± 1.5 d |
(2) MPE
(b) When a terminal interval is bounded by a surface, the MPE for any distance beginning at this point is increased by the value c given in Table 18.
Accuracy Class | a (mm) | b (mm) | c (mm) |
I | 0.1 | 0.1 | 0.1 |
II | 0.3 | 0.2 | 0.2 |
III | 0.6 | 0.4 | 0.3 |
D – special class for dipping tapes(¹) Up to and including 30 m(²) |
1.5 | zero | Zero |
S – special class for tank strapping tapes For each 30 m length when the tape is supported on a flat surface |
1.5 | zero | Zero |
Length i of the interval | MPE or difference in millimetres according to accuracy class | ||
I | II | III | |
i ≤ 1 mm | 0.1 | 0.2 | 0.3 |
1 mm < i ≤ 1 cm | 0.2 | 0.4 | 0.6 |
(3) Materials
(4) Markings
The nominal value shall be marked on the measure. Millimetre scales shall be numbered every centimetre and measures with a scale interval greater than 2 cm shall have all scale marks numbered.
Capacity serving measures
53.
—(1) The requirement for the instrument to bear information in respect of its accuracy shall not apply.
(2) Definitions
In these Regulations—
(3) Reference Conditions
(4) MPE
Table 20
Line
Brim
Transfer of measures
< 100 ml
± 2 ml
- 0
+ 4 ml
≥ 100 ml
± 3 %
- 0
+ 6%
Serving measures
< 200 ml
± 5 %
- 0
+ 10%
≥ 200 ml
± 5 ml + 2.5%
- 0
+ 10 ml + 5%
(5) Materials
Capacity serving measures shall be made of material which is sufficiently rigid and dimensionally stable to maintain capacity within the MPE.
(6) Shape
(a) Transfer measures shall be designed so that a change of contents equal to the MPE causes a change in level of at least 2 mm at the brim or filling mark.
(b) Transfer measures shall be designed so that the complete discharge of the liquid being measured will not be impeded.
(7) Marking
(3) The critical change value is equal to one scale interval.
Length measuring instruments
56.
—(1) Characteristics of the product to be measured
(2) Operating conditions
Group | Range of K | Product |
I | 0 < K < 2 x 10-² N/m² | low stretchability |
II | 2 x10-² N/m² < K < 8 x 10-² N/m² | medium stretchability |
III | 8 x10-² N/m ² < K< 24 x 10-² N/m² | high stretchability |
IV | 24 x 10-² N/m² < K | very high stretchability |
(3) MPE
Accuracy class | MPE |
I | 0.125 % , but not less than 0.005 Lm |
II | 0.25 % , but not less than 0.01 Lm |
III | 0.5 % , but not less than 0.02 Lm |
(4) Other requirements
The instruments must ensure that the product is measured unstretched according to the intended stretchability for which the instrument is designed.
Area measuring instruments
57.
—(1) Operating conditions
(2) MPE
58.
Instrument
Other requirements
59.
—(1) Presentation of the product
In the case of pulling back or stopping the product, it should not be possible to have an error of measurement or the display must be blanked.
(2) Scale interval
Multi-dimensional measuring instruments
60.
—(1) Operating conditions
Scale interval (d) |
Minimum dimension (min) (lower limit) |
d ≤ 2 cm | 10 d |
2 cm < d ≤ 10 cm | 20 d |
10 cm < d | 50 d |
(2) MPE
(c) a minimum temperature range of 50 °C for the climatic environment;
(d) the nominal value of the AC voltage supply and/or the limits of DC supply.
MPE
64.
The MPEs are—
Note: The error of the gas meter is not taken into account.
Permissible effect of disturbances
65.
—(1) Electromagnetic immunity
(2) The effect of an electromagnetic disturbance on a volume conversion device shall be such that—
(3) After undergoing a disturbance, the volume conversion device shall—
(4) The critical change value is the quantity corresponding to half of the magnitude of the relevant MPE as defined at paragraph 64.
Suitability
66.
—(1) A volume conversion device powered from the mains (AC or DC) shall be provided with an emergency power supply device or other means to ensure, during a failure of the principal power source, that all measuring functions are safeguarded.
(2) A dedicated power source shall have a lifetime of at least five years. After 90% of its lifetime an appropriate warning shall be shown.
(3) An indicating device shall have a sufficient number of digits to ensure that the quantity passed during 8,000 hours at Qmax by the gas meter to which the volume conversion device is to be connected does not return the digits to their initial values.
(4) The volume conversion device shall be able to be installed to operate in any position declared by the manufacturer in its installation instruction.
(5) An electronic volume conversion device shall be capable of detecting when it is operating outside the operating range(s) stated by the manufacturer for parameters that are relevant for measurement accuracy. In such a case, the conversion device must stop integrating the converted quantity, and may totalise separately the converted quantity for the time it is operating outside the operating range(s).
(6) An electronic volume conversion device shall be capable of displaying all relevant data for the measurement without additional equipment.
Units
67.
Metered quantity shall be displayed in cubic metres or kilograms.
Base conditions for converted quantities
68.
The manufacturer shall specify the base conditions for converted quantities.
Putting into Use
69.
Any volume conversion device may be used in conjunction with any gas meter put into use in accordance with Annex MI-002 of the Directive.
6.
The impartiality of the body, its director and staff shall be guaranteed. The remuneration of the body shall not depend on the results of the tasks it carries out. The remuneration of the body's director and staff shall not depend on the number of tasks carried out or on the results of such tasks.
7.
The body shall satisfy the designating authority that it has adequate civil liability insurance.
8.
The body's director and staff shall be bound to observe professional secrecy with regard to all information obtained in the performance of their duties pursuant to these Regulations, except vis-à-vis the authority of the designating authority.
in accordance with the Annex applicable to the relevant conformity assessment procedure in respect of a measuring instrument.
(2) In determining such an application, the notified body—
(3) Where, in the opinion of the notified body, the measuring instrument to which an application relates is compliant with the essential requirements, it shall issue a certificate or notification which is in accordance with the requirements of paragraph 12.
(4) Where, in the opinion of the notified body, the measuring instrument to which an application relates is not compliant with the essential requirements, it shall issue a notice to the applicant in accordance with paragraph 15.
(5) Where a certificate or notification under sub-paragraph (3) is issued by a notified body other than the Secretary of State, it shall send a copy to the designating authority from which it obtained its designation under regulation 4.
Limitations on duties to exercise functions
10.
—(1) A notified body shall not accept an application for a certificate or notification in respect of a measuring instrument unless the application—
(2) A notified body shall not be required to determine an application for a certificate or notification where the manufacturer has not—
(3) A notified body shall not be required to carry out the functions referred to in regulation 4(4) (d) if—
Contractors
11.
—(1) A notified body may, in exercising its functions—
(2) But nothing in sub-paragraph (1) authorises a notified body to rely on the opinion of another person with regard to whether a measuring instrument is compliant with any of the essential requirements.
(3) Nothing in these Regulations shall preclude a person referred to in sub-paragraph (1)(a) or (1)(b) from charging any fee in respect of any work undertaken by him in pursuance of those sub-paragraphs.
Form of certificates or notifications
12.
—(1) A certificate or notification issued by a notified body shall be in writing and, in addition to the requirements provided for in the conformity assessment procedure, shall—
(c) be signed by or on behalf of the notified body and give the identification number of the notified body.;
(d) bear—
(e) give particulars of the relevant measuring instrument (where applicable, in relation to each variant) to which it relates sufficient to identify it, and shall state whether the measuring instrument to which it relates is a single item or a representative, or if it covers a number of variants of that instrument; and
(f) certify that the measuring instrument to which it relates is compliant with the essential requirements.
Conditions of certificates or notifications
13.
—(1) A certificate or notification may be unconditional or may be subject to such conditions as the notified body considers appropriate.
(2) Such conditions may include—
(3) The conditions imposed pursuant to sub-paragraph (1) may be varied in accordance with paragraph 15 by the notified body which issued the certificate or notification and such a variation may include the imposition of new conditions or the removal of conditions.
Withdrawal of certificates or notifications
14.
—(1) The notified body which issued the certificate or notification shall withdraw that certificate or notification in accordance with paragraph 15, if it appears that the measuring instrument to which it relates is not compliant with the essential requirements.
Procedure where a notified body is minded to refuse to give, or to vary or withdraw a certificate or approval
15.
—(1) Where a notified body is minded to—
it shall give to the applicant, or the person to whom the certificate or notification was given, a notice in writing—
(2) Where a notified body, having considered representations made to it under sub-paragraph (1) remains of the opinion that—
it shall inform the applicant, or the person to whom the certificate or notification was given, of that decision in writing and give that applicant or person information about the judicial remedies available to him.
(3) Where a notice is given under sub-paragraph (1) by a notified body other than the Secretary of State, it shall send a copy to the designating authority from which it obtained its designation under regulation 4.
3.
The technical documentation shall include insofar as relevant for assessment and identification of the type and/or instrument—
(j) the EC-type examination certificates or EC design examination certificates in respect of instruments containing parts identical to those in the design.
4.
The manufacturer shall specify where seals and markings have been applied.
5.
The manufacturer shall indicate the conditions for compatibility with interfaces and sub-assemblies, where relevant.
[3] OJ No. L135, 30.4.04, p.1.back
[4] OJ No. L204, 21.7.98, p.37.back
[5] OJ No. L217, 5.8.98, p.18.back
[6] The application of Directive 2004/22/EC (OJ No. L135, 30.4.04, p.1 was extended to the European Economic Area by Decision No. 31/205 (OJ No. L198, 28.7.05, p.20).back
[17] OJ L 316, 31.10.92, p. 12. Directive abolished by Directive 2003/96/EC (OJ L 283, 31.10.03, p.51).back
[18] OJ L 42, 15.2.75, p.1. Directive as last amended by Directive 89/676/EEC (O) L 398, 30.12.89, p. 18).back
[19] OJ L 46, 21.2.76, p.1. Directive as last amended by the EEA Agreement.back
[20] OJ L220, 30.8.93, p.23.back